Category: Transient Plane Source
Author(s): S. Rudtsch, U. Hammerschmidt
Keywords: density, ghf, ghp, guarded heat-flow meter, guarded heat-flow meter (ghf), guarded hot-plate, guarded hot-plate (ghp) method, intercomparison of methods, method, pmma, poly(methyl methacrylate), polymethyl methacrylate (pmma), specific heat, technique, temperature, thermal conductivity, thermal contact resistances, thermal diffusivity, ths, tps technique, transient hot strip, transient hot strip (ths) technique, transient plane source, transient plane source (tps) method
Abstract: The authors used a variety of techniques to determine the thermal properties of polymethyl methacrylate (PMMA), including thermal conductivity, thermal diffusivity, specific heat and density. The measurements are performed in the temperature range from -70°C to 80°C. The results from the different techniques used to characterize the PMMA (GHP, GHF, TPS, THS) are all compared to see whether they agree with one another.
Reference: International Journal of Thermophysics, 25, 5 (2004) 1475-1482
DOI: 10.1007/s10765-004-5752-y