This paper introduces a compensation algorithm and a different-layers calibration method that will improve the temperature measurement speed and thermal resistance elimination of a guarded heat flow meter instrument. Without these additions to the method, heat will travel through a thin slice of material too fast for thermal couples to measure accurately. Thermal contact resistance would cause inaccuracy with a thin slice ...
The variability within the results presented from the guarded hot-plate (GHP) and guarded heat-flow meter (GHF) methods was determined. Multiple thermal properties of polymethyl methacrylate (PMMA) were analyzed to compare methods. These measurements are taken in the range of −70◦C and +80◦C. Providing reference data for PMMA is the main objective of this paper as well as quantifying the effects of thermal contact of the measuring instruments....
In this paper, a new method is proposed for measuring the through plane thermal conductivity of two catalyst layers of different thicknesses by the guarded heat flow meter (GHF) method, and the transient plane source method (TPS). These catalyst layers were made from two catalyst-coated substrates. The effects of pressing, and hot pressing the layers were also studied and shown to have ...