Discover the Next Chapter of Thermal Conductivity at ITCC-ITES 2024
Intracavity transducers are probes used in early pregnancy scans. To ensure the transducers do not exceed the maximum temperature threshold when in close contact with embryonic and fetal tissues, a thermal test object (TTO) was designed to measure the temperature of the transducers when in contact with soft tissue. A tissue mimicking material (TMM) was fabricated in place of real human tissue and coupled with the transducer. A Thermal Constants ...